bjr supplement 25 bjr supplement 25

The PDDs and dose profiles were measured using 2, 5, and 15 cm cone sizes at 80, 120, 140, and 150 kVp energies in a scanning water phantom using IBA Farmer-type ionization chambers of volumes 0.65 and 0.13 cc, respectively.

BJR SUPPLEMENT 25 SOFTWARE

This beam characterization and resulting source model was used as input for our in-house dose calculation software (kVDoseCalc) to compute radiation dose at points of interest (POIs). In-air ionization chamber measurements were made along the x- and y-axes of the X-ray beam to derive the spatial fluence and half-value layer (HVL) measurements were made to derive the spatially varying spectra. The Xstrahl source was modeled as an effective point source with varying spatial fluence and spectra. We also compared the measured and calculated PDDs to those in the BJR 25. We measured central axis PDDs and dose profiles using an Xstrahl 150 X-ray system. The purpose of this study was to empirically characterize and validate a kilovoltage (kV) X-ray beam source model of a superficial X-ray unit for relative dose calculations in water and assess the accuracy of the British Journal of Radiology Supplement 25 (BJR 25) percentage depth dose (PDD) data.






Bjr supplement 25